KANGAROO ePump
32
Enteral Feeding Pump
Feeding Formula VTBD
1-3000 mL in 1 mL increments
Bolus Volume
1-3000 mL in 1 mL increments
Number of Boluses
1-99
Bolus Interval
1-24 hours in 1-hour increments
Flushing Solution Dose Range
10-500 mL in 1 mL increments
Flushing Solution Interval Range
1-24 hr in 1 hr increments
Accuracy
±10% or 0.5 mL/hr, whichever is larger, with bag at 46 cm (18”) above pump, at room temperature 22° C ± 2° C (72° F ± 3° F), using
a new pump set with less than the recommended 24 hours of maximum usage.
Occlusion Pressure
15 psi (103 kPa) Nominal
Dimensions
Height: 16.8 cm (6.6”) Width: 16.3 cm (6.4”) Depth: 11.7 cm (4.6”)
Weight
1.1 kg (2.4 lbs), 1.2 kg (2.7 lbs) with pole clamp
Material
Housing: ABS/Polycarbonate blend
Pump Set Door: Polyester/ Polycarbonate blend
Power
Use AC adapter for wall outlet usage. Pump operates on 9V DC, 1.5 Amps.
Battery
New, fully charged NiMH Battery Pack delivers ≥ 15 hrs at 100 mL/hr feed rate. Approximately 15 minutes prior to complete battery
discharge, a low battery alarm will occur (see subsection “Battery Low” in Section IX). When complete discharge occurs, the pump
will automatically turn itself off.
The battery will charge continuously whenever the pump is plugged into a wall outlet. Eight hours of charging is required to fully
recharge the battery pack.
Alarms
System Error
Flow Error
Pump Set Dislodged Error
Hold Error
Rotor Error