-17-
4.8 Installation of the oil mist trap (Option)
An oil mist trap can be installed in order to remove oil mist from the pump. In G-100D,G-100S,
it is an OMT
-100A type. Remove the standard outlet pipe from the outlet port of the pump and
install an oil mist trap instead. The oil mist trap not only prevents oil mist generation but also
reduces exhaust noise by half.
For details, refer to the instruction manual for the oil mist trap.
4.9 Restriction on operation when the oil mist trap is installed
When using the oil mist trap, there are some restrictions on operation. When the filter is clogged,
replace it.
The internal pump pressure is limited to 0.03 MPa (gauge pressure). When the pressure
measured at the outlet side has reached 0.03 MPa (gauge pressure), replace the oil mist trap
filter.
Be sure to observe the restrictions on operation when the oil mist trap is
installed. There is a risk of explosion. When the filter is clogged, replace
it.
Warning