UT219PV User Manual
14 / 27
ACV-LPF
Warning:
Do not input voltage over 2500V DC or 1500V AC. It is possible to measure higher voltage, but it may
cause damage to the Clamp Meter.
● For high voltage measurement, please pay special attention to avoid electric shock.
If the measured voltage is ≥30V (AC/DC), the high-voltage warning symbol will be displayed.
“OL” will appear on the LCD if the measured voltage is >2510V DC and <-2510V DC or >1510V AC.
If the DC voltage is less than -10V, the red backlight is lit up and the LED blinks for 10 seconds, the
buzzer sounds for 10 seconds, and the symbol “POLARITY” blinks.
If frequency is displayed on the sub-display, then the input amplitude shall be ≥5V rms for
20Hz~100Hz and ≥10V rms for 100Hz~1000Hz.
The low pass filter attenuates at -3dB and the cutoff frequency is 1.2 KHz.
3. Measuring DC power and DC voltage/current (Figure 7)
Figure 7