UT-1401
DP INDICATOR AND SENSOR
10
‘Calibration Mode’ Mode will shown by pressing key, after Mode display.Setp
4.3 Calibration Mode
Range / Values
Parameter
CAL
0000
CALIBRATION MODE
Password for Calibration Mode Parameters
Display
Scroll the digit by key, Increase by key.
Range : Values between to0000 9999
Default Value : 0000
CALz
----
CALIBRATION ZERO
Calibration Zero
NOTE : AfterZeroCal.press keyfornextPara.
Apply Low signal [as per input] & press key.
Range :
Default Value : ----
According to Input Type
CALS
----
CALIBRATION SPAN
Calibration Span
NOTE : After Span Cal. press key for next para.
Apply High signal [as per input] & press key.
CALL
strt
CALIBRATION LOW OUTPUT
Calibration desired Low Output at PV Display.
NOTE : After out low cal. press key for next Para.
Set output by , & key, press key
for desired low measured output achieved.
Note : After Calibration random value
appeared in display, as per input.
Refer below Table for Cal. values range.
Note : After Calibration random value
appeared in display, as per input.
Refer below Table for Cal. values range.
Range :
Default Value : ----
According to Input Type
Note : After Cali. random value should be
below appeared in display.1000
Range :
Default Value : -> to start cal.strt
According to Output Type
CALH
strt
CALIBRATION HIGH OUTPUT
Calibration desired High Output at PV Display.
NOTE : After out low cal. press key for next Para.
Set output by , & key, press key
for desired high measured output achieved.
Range :
Default Value : -> to start cal.strt
According to Output Type
Note : After Cali. random value should be
below appeared in display.8000