Check Is there anything obstructing the platter?
Action Remove all obstacles.
Syndrome Weighing result is not accurate
Check Is the scale placed in a level condition?
Action Adjust the adjustable feet to a level
condition.
Check Is the scale affected by airflow, vibration
or RFI?
Action: Place the scale away from all disturbances.
Check Is the scale calibrated correctly?
Action Contact your dealer.
Check Is the correct weight unit employed?
Action Refer to 5.3 for weight unit setting and
usage.
Syndrome Display shows 00000 after counting down
Check Is the scale overloaded?
Action Remove some loads from platter.
Check Is the stainless steel platter cover placed
on the platter?
Action Place the stainless steel platter cover
onto the platter.
7. DAILY CARE AND MAINTENANCE
7.1 Clean the scale with a soft and damp cloth. If
necessary, apply a mild detergent.
7.2 Do not use any harsh, abrasive material,
acetone, volatile solvent, thinner or alcohol
for cleaning.
7.3 Verify the accuracy of scale periodically.
Re-calibrate the scale if necessary.
NOTE: In some countries, calibration is restricted to
be done by an authorized/qualified agent only. Contact
your dealer for more information.
7.4 It is a good practice to apply the dust cover
when operating the scale.
7.5 The scale must be placed horizontally during
transportation or long time storage.
7.6 Remove platter from scale before transportation
or long time storage.
7.7 Store scale in a dry and clean place.