Seite 1 von 72 // page 1 of 72
Das Dokument darf nur vollständig und unverändert verwendet und weitergegeben werden. Es liegt in der Verantwortung des
Anwenders, die Gültigkeit dieses Dokumentes bezüglich seines Produktes sicher zu stellen.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product.
Instandsetzungsanleitung
Instructions for repair
Drehschieberpumpen / Rotary vane pumps
RE 2 / RZ 2 / RE 2.5 / RZ 2.5 / RE 5 / RZ 5 / RE 6 / RZ 6 /
RE 8 / RZ 8 / RE 9 / RZ 9 / RE 16 / RZ 16
Chemie-HYBRID-Pumpen / Chemistry-HYBRID-Pumps
RC 5 / RC 6
Vakuum-Messgerät / Vacuum gauge
VAP 5
Vakuumtechnik im System
Technology for Vacuum Systems
Instandsetzungsanleitung: 999111 - 07 / Stand 01/07/2018
Gültig für Drehschieberpumpen Typ RE 2, RZ 2, RE 2.5, RZ 2.5, RE 5, RZ 5, RE 6, RZ 6,
RE 8, RZ 8, RE 9, RZ 9, RE 16, RZ 16
HYBRID Pumpen Typ RC 5, RC 6
Vakuummessgerät VAP 5
Die Gültigkeit des pumpenspezifischen Zeichnungs- und Ersatzteileblatts ist anhand der Serien-
nummer der Pumpe vom Anwender zu prüfen.
Für Pumpen älterer oder neuerer Ausführung bitte das pumpenspezifische Zeichnungs- und Er-
satzteileblatt unter Angabe von Pumpentyp, Seriennummer und Baujahr bei VACUUBRAND anfor-
dern.
Instructions for repair: 999111 - 07 / version 01/07/2018
Valid for Rotary vane pumps type RE 2, RZ 2, RE 2.5, RZ 2.5, RE 5, RZ 5, RE 6, RZ 6,
RE 8, RZ 8, RE 9, RZ 9, RE 16, RZ 16
HYBRID pumps type RC 5, RC 6
Vacuum gauge VAP 5
It is the users´ responsibility to check the validity of the pump specific drawing and spare parts list on
the basis of the serial number of the pump.
If the pump is older or newer please contact VACUUBRAND and order the pump specific drawing
and spare parts list stating the pump type, serial number and year of manufacture.

Questions and Answers

  • M
    Mackenzie JacksonAug 10, 2025
    Why is my vacuubrand RE 2 display alternating on and off?
    • K
      Kenneth HardinAug 10, 2025
      The display might be alternating on and off because the internal overload protection is activated due to thermal overload. Ensure adequate ventilation to lower the ambient temperature. If the issue persists, the device may be defective and require factory repair.
  • G
    greenenicholasAug 14, 2025
    What to do if my vacuubrand RE 2 pressure reading is fluctuating?
    • K
      Kyle RangelAug 15, 2025
      Fluctuating pressure readings can occur due to several reasons: * Pressure fluctuations in the vacuum system itself. Check and adjust the system setup as needed. * Loose gauge head plug. Ensure the plug is securely screwed into the female connector. * Interface interference at the DVR 5. Eliminate the source of interference or use an interface filter. * Change in position of the gauge head VSK 5 (e.g., horizontal to vertical). Readjust if precise pressure indication is necessary.
  • S
    Steven WilliamsonAug 20, 2025
    What to do if my vacuubrand RE 2 Measuring Instruments shows no display?
    • M
      Melissa CareyAug 20, 2025
      If there is no display, first check if the mains plug is properly inserted and if there is power available at the socket. Also, check the mains fuses. If these steps don't resolve the issue, the device might be defective and require factory repair.
  • M
    Mr. Phillip CurtisAug 25, 2025
    What to do if my vacuubrand RE 2 pressure reading is too high in the low pressure range?
    • N
      Nicole CardenasAug 25, 2025
      If the pressure reading is too high in the low-pressure range, it could be due to: * A clogged protective filter in the gauge head VSP 5. Clean or replace the filter. * A contaminated gauge head. Clean the gauge head and readjust it. * Incorrect adjustment. Readjust the gauge head. * Poor electrical contact. Clean the socket pins of the cable.
  • M
    Marilyn ScottAug 29, 2025
    Why is the malfunction indicator flashing on my vacuubrand RE 2 Measuring Instruments?
    • A
      Ann AndersenAug 29, 2025
      If the warning triangle is flashing and "Err" is displayed, it indicates a malfunction with the pressure transducer. Ensure the gauge head is properly connected. If the problem persists, the gauge head or connecting cable may be defective and need replacement.
  • A
    Amber George DDSSep 4, 2025
    How to adjust vacuubrand Measuring Instruments if pressure reading is too low at atmospheric pressure?
    • D
      Dr. Erica WillisSep 4, 2025
      If the pressure reading is too low at atmospheric pressure, readjust the gauge head. If you are using VAP 5 and gases other than air or nitrogen, readjust the gauge head at atmospheric pressure with the specific gas being used.

vacuubrand RE 2 Specifications

General IconGeneral
Brandvacuubrand
ModelRE 2
CategoryMeasuring Instruments
LanguageEnglish