34
Reminder
Use the
Reminder
option to set a
reminder for calibrating the probe.
You can choose a preferred
calibration interval of 3, 6, 12, or 24
months. Calibrating the probe
resets the calibration interval. To
remove the reminder from use,
select
Disabled
.
Note that you must set the correct
date in the indicator for the
calibration reminder to appear as
intended.
The calibration reminder appears
one month before the preferred
calibration date is due.
To continue displaying the
reminder for this interval, select
Snooze
. The reminder will appear
again at each power-on.
To hide the reminder, select
Close
.
If you do not calibrate within a
month, the reminder will appear
again.
Calibrating any parameter in the probe resets
the calibration interval. This means that if you
leave one parameter uncalibrated in a
calibration session, the reminder will not
appear separately for that parameter.