6.10.9 Close the connection in POLL mode
The CLOSE command closes the connection to the probe.
CLOSE<cr>
Example:
close
line closed
6.10.10 Reset the probe
Use the RESET command to reset the probe. Upon reset or power-up, the probe enters the
serial mode that has been set with the SMODE command.
RESET<cr>
Example (probe set to serial mode STOP, will output probe model and software version at
reset):
reset
HMP60 / 1.00.0
Example (probe set to serial mode RUN, will start to output measurement messages at reset):
reset
T= 23.6 'C RH= 20.2 %RH Td= -0.5 'C
T= 23.6 'C RH= 20.2 %RH Td= -0.5 'C
T= 23.3 'C RH= 20.2 %RH Td= -0.8 'C
...
6.10.11 Restore factory settings
Use the FRESTORE command to restore the factory settings to the probe. All user settings,
including the user-performed calibration corrections, will be lost. The probe will revert back to
the factory calibrated settings.
FRESTORE<cr>
Example:
frestore
Factory settings restored
HMP60 and HMP110 Series User Guide M211060EN-J