3. Pull out the old INTERCAPâ sensor and insert a new one. Handle the new sensor by the
plastic frame.
Do not touch the sensor plate.CAUTION!
4. Attach a new filter on the probe.
7.5.2 Changing the HUMICAP 180R/180V sensor (HMP110 and
HMP113)
Replacing the humidity sensor of the HMP110 and HMP113 is not necessary in
normal operation. If the accuracy of the probe does not seem to be within
specification, it is likely that the accuracy can be restored by performing the
adjustment procedure.
Follow this procedure to replace the humidity sensor of the HMP110 and HMP113 in case it has
been damaged, or normal adjustment is not sucient to restore the measurement accuracy.
Calibration and adjustment of the humidity measurement is required after the sensor change.
To perform this procedure, you need a new HUMICAPâ 180R/180V sensor. It is also
recommended that you replace the filter with a new one.
1. Remove the filter from the probe by turning it counter-clockwise.
HMP60 and HMP110 Series User Guide M211060EN-J