EasyManuals Logo

Vaisala HMP60 series User Manual

Vaisala HMP60 series
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #73 background imageLoading...
Page #73 background image
Table 18 HMP60 calculated parameters
Property Description/Value
Measurement ranges
Dew point temperature −40 … +60 °C (−40 … +140 °F)
Wet bulb temperature −40 … +60 °C (−40 … +140 °F)
Absolute humidity
0 … 130 g/m
3
(0 … 57 gr/ft
3
)
Mixing ratio 0 … 153 g/kg (0 … 1069 gr/lb)
Enthalpy −40 … 459 kJ/kg (−9.6 … 207 BTU/lb)
Table 19 HMP60 operating environment
Property Description/Value
Operating temperature −40 … +60 °C (−40 … +140 °F)
EMC compliance EN 61326-1, industrial environment
Table 20 HMP60 inputs and outputs
Property Description/Value
Power consumption 1 mA average, max. peak 5 mA
Operating voltage
1)
With 1 V / 2.5 V output 5 … 28 VDC
With 5 V output 8 … 28 VDC
With loop power converter 8 … 28 VDC
With digital output 5 … 28 VDC
Start-up time
Probes with analog output 4 s at operating voltage 13.5 … 16.5 VDC
2 s at other valid operating voltages
Probes with digital output 1 s
Outputs
2 channels 0 … 1 VDC / 0 … 2.5 VDC / 0 … 5 VDC / 1 … 5 VDC
1-channel loop-power converter (separate
module, compatible with humidity accuracy
only)
4 … 20 mA
Digital output (optional) RS-485 2-wire half duplex, supports Modbus
RTU
Chapter 9 – Technical data
71

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala HMP60 series and is the answer not in the manual?

Vaisala HMP60 series Specifications

General IconGeneral
BrandVaisala
ModelHMP60 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals