EasyManua.ls Logo

Vaisala HMP60 series

Vaisala HMP60 series
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Property Description/Value
at −40 … 0 °C and +40 … +60 °C (−40 … +32 °F
and +104 … +140 °F)
±5 %RH (0 … 90 %RH)
±7 %RH (90 … 100 %RH)
Humidity sensor Vaisala INTERCAPâ
Temperature
Measurement range −40 … +60 °C (−40 … +140 °F)
Accuracy:
at +10 … +30 °C (+50 … +86 °F)
at −40 … +10 and +30 … +60 °C (−40 … +50 and
+86 … +140 °F)
±0.5 °C (±32.9 °F)
±0.6 °C (±33.08 °F)
Dew point temperature
Measurement range −40 … +60 °C (−40 … +140 °F)
Typical accuracy:
at 0 … +40 °C (+32 … +140 °F) when dew point
depression
1
< 15 °C
±2 °C (±35.6 °F)
at −40 … 0 °C and +40 … +60 °C (−40 … +32 °F
and +104 … +140 °F) when dew point depression
< 10 °C
1)
±3 °C (±37.4 °F)
Analog outputs
Accuracy at 20 °C (+68 °F) ±0.2 % of FS
Temperature dependence ±0.01 % of FS/°C (±0.006 % of FS/°F)
1) Dew point depression = ambient temperature − dew point
Table 23 HMP63 calculated parameters
Measurement ranges
Dew point temperature −40 … +60 °C (−40 … +140 °F)
Wet bulb temperature −40 … +60°C (−40 … +140°F)
Absolute humidity
0 … 130 g/m
3
(0 … 57 gr/ft
3
)
Mixing ratio 0 … 153 g/kg (0 … 1069 gr/lb)
Enthalpy −40 … 459 kJ/kg (−9.6 … 207 BTU/lb)
Chapter 9 – Technical data
73

Table of Contents