Chapter 3 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 25
Preparing the Sounding
When preparing the sounding, Vaisala Ground Check Set GC25 is
connected to the sounding system via cable and operated with the help
of the sounding software.
0705-025
Figure 13 Vaisala Ground Check Set GC25
Follow these steps to prepare the sounding:
1. Open the chamber of the Ground Check Set and place the
radiosonde onto the radiosonde tray. Close the chamber.
NOTE
If you are using DigiCORA® Sounding System MW21, software
version < 3.12, or MW15, MW12 or MW11, software version < 8.311,
refer to the Ground Check Set GC25 User’s Guide for information on
using the GC25 in the stand-alone mode.
1 = Chamber
2=Display
3 = Buttons
4 = Communication cable
5 = Radiosonde tray
6=Power switch