EasyManua.ls Logo

Vapotherm precision flow - System Overview and Features; Key Features of Precision Flow

Vapotherm precision flow
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 6 3001002 Rev. P
Section 2 Overview
The Precision Flow® is a system for high flow humidified respiratory therapy by nasal
cannula. It incorporates the Vapotherm core humidification technology with an electronic
blender and flow controller. The water and gas pathways are both incorporated into a
removable, disposable patient circuit.
Features
The patient circuit is detachable and disposable: no
disinfection necessary
Minimal downtime between patients: less than
five minutes to change disposables
Built-in oxygen/air blender
Built-in electronic flowmeters and controllers
Self-testing and self-calibrating
Internal battery backup maintains flow and
oxygen percentage for at least 15 minutes if AC
power is cut off. Battery recharges in 2 hrs.
All internal sensors self-calibrating and self
monitoring
Single button starts and stops the device
Temperature, flow and oxygen percentage are
adjusted via a single setting control knob on
the front panel
All values and alarms displayed in a single large
color-coded panel
Flow range 1-40 lpm
Oxygen percentage is fully adjustable from 21 to 100% when two 40 psi (276 kPa) gas
sources are used
Inlet gas pressure range is 4-85 psi (28-586 kPa)
Single gas operation- the Precision Flow
®
detects inlet gas pressure and blends flow
based on demand required and available supply. Supply pressure determines FiO
2
and
delivered flow; if demand exceeds supply an alarm sounds
At low gas inlet pressures, maximum flow rate and oxygen percentage settings are
automatically reduced to match the inlet pressure
Automatically senses cartridge type: maximum flow setting is automatically reduced if
low-flow cartridge is installed
Warm-up time less than five minutes
Sterile water is connected to the disposable water path using a standard
spike
Universal power requirements allow use anywhere with only a change of power cord
Scheduled maintenance: gas trap filters replaced at 6-month intervals, and oxygen
sensor replaced annually
Precision Flow®