8. Product Specifications
EzRay Air W (Model: VEX-S300W) User Manual 57
ENGLISH
To assure line voltage quality, a separate 3-core grounded power
cable connected directly to central distribution panel with over-
current circuit breaker rated for 10A must be used.
The mains resistance should not exceed 0.5 Ω.
The system will be available with a fixed tube current specification
based on the user selection.
Power Supply is specified as a part of ME EQUIPMENT.
Power plugs may have various specifications for each country.
8.4 Environmental Specifications
Item Description
During operating
Temperature
10 ~ 35 ℃
Relative humidity 30 ~ 75 %
Atmospheric pressure 860 ~ 1060 hPa
Transport and storage
Temperature
-10 ~ 60 ℃
Relative humidity 10 ~ 75 % non-condensing
Atmospheric pressure 860 ~ 1060 hPa