‘U’ - Followed by 4 to 5 characters.
Direct Flow Gain (G3).
(e.g. 2.00 = 2.00 x gain)
(e.g. 10.00 = 10.00 x gain)
In software revisions 3.8 and later, this field is fixed at 5 characters.
(‘U’ + 4 numbers).
(e.g. 0971 = 9.71 x gain)
‘V’ - Followed by 1 to 4 numbers.
Crossover #1 or Reverse Flow Crossover in millitorr.
(e.g. 10 = 10 millitorr)
(e.g. 200 = 200 millitorr)
(e.g. 9000 = 9000 millitorr)
In software revisions 3.8 and later, this field is fixed at 5 characters.
(‘V’ + 4 numbers).
(e.g. 0300 = 300 millitorr)
‘W’ - Followed by 4 or 5 characters.
Stamped Leak Rate on Internal Calibrator in atm-cc/sec when state of
leak detector is in any state except I-47. When the leak detector is
in state I-47, this field will be the temperature compensated value of the
internal calibrator. This value is what the leak detector is calibrating to
when the internal calibrator is selected as the active calibrator.
(e.g. 5708 = 5.7 x 10-8 atm-cc/sec)
(e.g. 10010 = 10.0 x 10-10 atm-cc/sec)
In software revisions 3.8 and later, this field is fixed at 5 characters.
(‘W’ + 4 numbers).
(e.g. 6304 = 6.3 x 10-4 atm-cc/sec. A mantissa of 10.0 will never be
sent.)
‘X’ - Followed by 4 characters.
Stamped Temperature on Internal Calibrator in Degrees Celsius.
(e.g. 20.2 = 20.2 °C)
(e.g. 35.4 = 35.4 °C)
In software revisions 3.8 and later, this field is fixed at 3
characters.
(‘X’ + 2 numbers).
(e.g. 25 = 25°C)
1109-186-00 Rev. C d-viii