EasyManua.ls Logo

Vega VEGAMET 624 - Flow Measurement

Vega VEGAMET 624
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
l Select under "Reaction in case of failure" the option "Switching
condition o"
l Enter the following values into the parameter elds:
- Measured value more than 25 %/min.
- Max. reaction time 1 min.
The function of the tendency recognition is shown in detail in the
following diagram. The previously described example is used as a
basis.
t
t
m
t
m
t
m
t
m
120 180 240
ON
OFF
60
[sec]
1 2 3 4
5
...
...
...
...
100
75
50
25
0
%
...
Fig. 12: Example for tendency recognition
1 Old average value = 25 %, new average value = 25 %
Dierence < 25 % -> Relay ON
2 Old average value = 25 %, new average value = 37.5 %
Dierence < 25 % -> Relay ON
3 Old average value = 37.5 %, new average value = 62.5 %
Dierence = 25 % -> Relay OFF
4 Old average value = 62.5 %, new average value = 75 %
Dierence < 25 % -> Relay ON
5 tm -> max. reaction time
8.4 Flow measurement
For ow measurement in open umes, a contraction or standardised
ume must be used. Depending on the ow volume, this contraction
generates a certain backwater. The ow can be determined from the
height of this backwater. The ow volume is outputted by an
appropriate number of pulses on the relay or current output.
Depending on the type and version, each ume generates a dierent
backwater. The data of the following umes are available in the
instrument:
l Palmer-Bowlus-Flume
l Venturi ume, trapezoidal weir, rectangular weir
l V-Notch
Setup
The conguration of the ow measurement loop requires PACTware
with the suitable DTMs. The example refers to a ow measurement
with a radar sensor. The following setup steps must be carried out:
Functional principle
Flume
40 VEGAMET 624 4 20 mA/HART signal conditioning instrument
8 Application examples
28969-EN-091116

Other manuals for Vega VEGAMET 624