l [+] key:
- Change value of the parameter
l [ESC] key:
- interrupt input
- jump to the next higher menu
Note:
Approx. 10 minutes after the last pressing of a key, an automatic reset
to measured value indication is triggered. Any values not conrmed
with [OK] will not be saved.
6.2 Setup procedure
Through parameter adjustment, the instrument is adapted to the
individual application conditions. Calibration of the measuring point is
the most important step and should always be carried out. A scaling of
the measured value to the desired physical variable and unit, possibly
including a linearization curve, is often useful. The adaptation of the
relay switching points or the setting of an integration time to smooth the
measured value are further standard adjustment options.
For instruments with Ethernet interface, the instrument must be
provided with an IP address and subnet mask suitable for the
instrument. As an alternative, addressing via DHCP and a host name
is also possible. If necessary, the e -mail/webserver can be also
congured with PACTware.
Information:
When using PACTware and the corresponding VEGA DTM, additional
settings can be carried out which are not or only partly possible with
the integrated indicating and adjustment unit. When using an adjust-
ment software, you either need one of the integrated interfaces
(RS232/Ethernet) or the interface converter VEGACONNECT.
Further instructions for setting up the web server and e-mail functions
are stated in the online help of PACTware or the VEGAMET 624 DTMs
as well as the supplementary instructions manual "RS232/Ethernet
connection".
After being switched on, VEGAMET 624 rst of all carries out a short
self-check. The following steps are carried out:
l Internal check of the electronics
l indication of the instrument type, rmware version as well as the
instrument TAG (instrument name)
l The output signals jump briey to the set fault value
Then the current measured values will be displayed and outputted.
Parameter adjustment
Switch on phase
VEGAMET 624 4 20 mA/HART signal conditioning instrument 17
6 Setup with the integrated indicating and adjustment unit
28969-EN-091116