HPS10 PersonalScope
TM
Trigger function
Press the ‘t-V/div’-key to set the cursorkeys
action into ‘timebase’ mode.
Press the ‘left’ or ‘right’ cursor keys to increase
or decrease the timebase (time/div).
Note:
Changing the timebase or sensitivity
switches the autorange mode immediately
off. (Fig 37)
The timebase or sensitivity cannot be
changed into hold mode.
Pressing the ‘t-V/div’ key into hold mode
toggles the screen between the two stored
waveforms.
At higher timebase (2us and faster) the scope
uses oversampling mode, only repetitive
signals are correctly displayed.
Use the minimum time base (0.2 microsec)
as a starting point when measuring a signal
and select longer time bases until the signal
is displayed properly. Otherwise the display
may not correctly reflect the signal under
measurement due to aliasing.
THE TRIGGER SETUP
Press first the trigger“-key.
Use the left cursorkey to change the triggermode
(norm, run, once or roll).
Use the right cursorkey to toggle the trigger slope.
Use the up/down cursorkeys to shift the vertical
position of the signal.
trig
Fig 37
HPS10_man.pub
page 46
Wednesday, April 03, 2002 09:22