User Guide
VentMed
light industry environments
Class II (double insulation),
IEC
60601-1:2005+A1: Type BF, Ingress protection
2012 classification IPX1.
Sensors
Pressure sensor
Internally located at device
outlet, analog gauge
pressure type, 0 to 40 cm
H2O
Flow sensor
Internally located at device
outlet, analog gauge flow
type, 50 to +200 L/min
Maximum single
fault steady
pressure
Device will shut down in
the presence of a single fault if
the steady state pressure
exceeds: 35 cm H2O for more
than 1 sec.
Sound
Pressure level measured
according to ISO 80601-2-70
(CPAP mode)Declared
53 dBA with uncertainty of
2 dBA under background
sound level 36 dBA
dual-number
noiseemission A-weighted sound power level
values in accordance with ISO is 61dba under background
4871:1996.
sound level 36dBA
280 mm x 140 mm x 95 mm
22 mm
Dimensions (H x W x D):
Air outlet (complies with ISO
5356-1:2004)
Physical-device
and water tank
Weight (device and water tank) 1.6kgs
Water capacity
To maximum fill line
220 ML, minimum 100ML
Injection molded plastic
and silicone seal
68°C
Water tank - material
Maximum heater plate
Cut-out
Temperature
102°C