performance and reduce the life of the products.
Periodically reassess the setting(s) of the therapy for effectiveness.
60W power supply AC input range
unit (Only use
100–240V~, 50–60Hz,
2A maximum, Class II
24V 2.5A
with IEC60601-1 DC output
approved adaptor)
+41°F to +95°F (+5°C to
+35°C)
Note: The air flow for breathing
produced by this therapy
device can be higher than the
temperature of the room.
Under extreme ambient
temperature conditions
(104°F/40°C) the device
remains safe.
Operating humidity:
Operating altitude
10 to 95% relative humidity,
non-condensing
Sea level to 8,500' (2,591 m);
air pressure range 1013 hPa to
738 hPa
Storage and transport
temperature
-4°F to +140°F (-20°C to +60°C)
Storage and transport humidity 5 to 95% relative humidity,
non-condensing
Electromagnetic
compatibility
The VentMed device complies
with all applicable
electromagnetic compatibility
requirements (EMC) according
to IEC60601-1-2:2014, for
residential, commercial and