Pressure sensor at air DS-5 CPAP
DS-7 BIPAP ST25: 4–25 0.1 cmH2O
DS-8 BIPAP ST30: 4-30
cmH2O
Static Airway Pressure ±0.5cmH2O + 4% of the set pressure
measurement: Mask
pressure
±3cmH2O + 5% of the set pressure
measurement(Both
Cpap/BIPAP): Mask
pressure
The percentage of each inspiratory and expiratory phase
to be taken into the calculation for determining the
accuracy is (I/E=1:1)
Remark: Results are expressed at ATPD (Ambient Temperature and Pressure,
Dry).