Displayed values
Value
Range
Display resolution
Pressure sensor at air DS-5 CPAP
0.1 cmH2O
0.1cmH2O
outlet: Mask pressure
DS-6 Auto CPAP
DS-7 BIPAP ST25: 425 0.1 cmH2O
cmH20
0.1 cmH2O
DS-8 BIPAP ST30: 4-30
cmH2O
Value Accuracy
Value
Accuracy
Static Airway Pressure ±0.5cmH2O + 4% of the set pressure
measurement: Mask
pressure
Dynamic Airway
Pressure
±3cmH2O + 5% of the set pressure
measurement(Both
Cpap/BIPAP): Mask
pressure
The percentage of each inspiratory and expiratory phase
to be taken into the calculation for determining the
accuracy is (I/E=1:1)
Remark: Results are expressed at ATPD (Ambient Temperature and Pressure,
Dry).
Environmental information
This device should be disposed of separately, not as unsorted municipal waste. To
dispose of your device, you should use appropriate collection, reuse and recycling
systems available in your region. The use of these collection, reuse and recycling
systems is designed to reduce pressure on natural resources and prevent hazardous
substances from damaging the environment.
If you need information on these disposal systems, please contact your local waste
administration. The crossed-bin symbol invites you to use these disposal systems.
If you require information on collection and disposal of the device please contact
your care provider or VentMed.