User Guide
VentMed
Portable and mobile RF
communications equipment should
be used no closer to any part of the
device, including cables, than the
recommended separation distance
calculated from the equation
application to the frequency of the
transmitter.
Recommended separation distance
d=1.2 √p
Conducted
RF
d=1.2√p
d=2.3√p
80MHz to 800MHz
800MHz to 2.5GHz
3Vrms
3Vrms
IEC61000-4 150kHz to
-6
80MHz
Immunity test
IEC60601 test
Compliance
Electromagnetic
environment-guidance
Where P is the
level
level
Radiated RF
3V/m
3V/m
maximum output
power rating of the
IEC61000-4-3
80MHz to
36