How to fix Vericom Measuring Instruments when it doesn't turn on?
M
Mandy MartinAug 15, 2025
If your Vericom Measuring Instruments unit isn't turning on, there might be several reasons. First, check the connection to the cigarette lighter plug, the fuse in the plug, and the power to the cigarette lighter. Second, make sure the cigarette lighter cord is fully connected to the VC4000 by pushing the right-angle plug all the way in. Finally, if the battery is dead, plug the unit into the cigarette lighter or a wall adapter to charge it.
J
John HammondAug 18, 2025
What to do if Vericom Measuring Instruments turns off during run?
A
Amy LewisAug 18, 2025
If your Vericom Measuring Instruments unit turns off or resets during operation, there could be a few reasons. First, check the power cord and fuse to ensure there is no loss of power to the VC4000. Second, the battery might be dead, so recharge it. Third, electrical noise interference could be the cause, so try to eliminate the noise from its source.
H
Heather GomezAug 25, 2025
What to do if Vericom VC4000 Measuring Instruments is unresponsive to keypad?
J
Joseph GriffinAug 26, 2025
If your Vericom Measuring Instruments unit is unresponsive to the keypad, it could be due to a program lock-up. To reset it, short pins 1 and 2 of the external activation input. Also, check to see if updated firmware is available.
L
Lucas GonzalezAug 28, 2025
What to do if Vericom VC4000 Measuring Instruments power is on but display is blank?
L
lowelisaAug 28, 2025
If your Vericom Measuring Instruments display is blank but the power is on, it might be due to sudden voltage swings. Try pressing the CLEAR / NO key on the keypad.
G
Gabrielle AnthonySep 1, 2025
How to fix Vericom Measuring Instruments GPS when it's not working?
M
Ms. Julie Gutierrez MDSep 1, 2025
If the GPS is not working on your Vericom Measuring Instruments (PC), first ensure that GPS speed is turned on in the Setup. Next, verify that the GPS RS232 port is set to 1. Finally, allow up to one minute for the system to acquire satellites.
A
Anthony CarrSep 4, 2025
What does FLASH-ERROR mean on Vericom VC4000?
M
Marilyn PerkinsSep 4, 2025
If your Vericom Measuring Instruments shows a FLASH-ERROR, it indicates a problem writing to the flash memory. If necessary, import the data to Profile and then clear the memory.
C
craigfarmerSep 7, 2025
What does RUN-FILE-FULL mean on Vericom VC4000 Measuring Instruments?
D
Dr. Nathaniel BradshawSep 7, 2025
If your Vericom Measuring Instruments displays RUN-FILE-FULL, it means the memory is full. You should import data to Profile, if necessary, and then clear the memory.
D
Darlene MccannSep 12, 2025
What does ARITHMETIC EXCEPTION mean on Vericom Measuring Instruments?
J
Jennifer StevensSep 12, 2025
If your Vericom Measuring Instruments displays ARITHMETIC EXCEPTION, it's due to a divide-by-zero error. Power the unit off and then back on to resolve this issue.
C
Christina JohnsonNov 20, 2025
Why is my Vericom VC4000 Measuring Instruments showing inaccurate data?
V
Virginia LoganNov 20, 2025
If your Vericom Measuring Instruments unit is providing inaccurate data, consider these potential causes. First, ensure that the VC4000 is at the same incline as when it was zero adjusted, and that the level hasn't changed since. Refer to page 4 for mounting instructions. Also, make sure summation is turned on if the unit is mounted at an angle. A bump may have triggered the timing to start too soon if 3D summation is on; see page 65 to adjust the threshold. Electrical noise interference can also cause inaccurate data, so eliminate any noise from the source. Check page 67 to make sure the pitch and roll factor is correct for your vehicle. If the settle time was calculated incorrectly, import the data to Profile and set it manually. Vibration can also cause inaccurate readings; mount the un...