x The completion of the calibration procedure is signalled
by the following message:
x Take off the weight from the pan, the message <donE> is displayed
for 1s and the program returns to the calibration submenu:
x Calibration process can be terminated anytime by pressing
which is signalled by the following message on the display:
x Return to weighing with saving changes that have been made.
Caution:
If the calibration process (span adjustment) lasts longer than 15 the
<Err8> message will be displayed and short audible signal will be
emitted. Press
to perform calibration again with more stable
ambient conditions!
17.3. Start mass adjustment
Option only for non-verified scales
Not accessible in WLC…/C/2 scales of WLC series
The scale can require only adjusting new start mass. In this scale adjusting
start mass can be excluded from the process of calibration and performed
separately.