EasyManua.ls Logo

Victaulic FireLock 751 Series - Removing and Replacing the Clapper Seal; Clapper Seal Removal Steps

Victaulic FireLock 751 Series
30 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
I-751.Europe / European Alarm Check Valve Stations / Installation, Maintenance, and Testing Manual
REMOVING AND REPLACING THE CLAPPER
SEAL (ALL SIZES)
NOTICE
The clapper assembly for an 8-inch/DN200 valve is shown in the
photos featured in this section.
1.
Perform steps 1 – 6 of Section VI, Required Internal Inspection.
2. Remove the self-sealing screw from the clapper seal.
3. Remove the seal retaining ring. Save the seal retaining ring for
re-installation.
CAUTION
DO NOT pry the seal washer out of the clapper seal from the
inner hole (for the 8-inch/DN200 valve size only).
Failure to follow this instruction could damage the seal washer,
resulting in improper clapper sealing and valve leakage.
4. For the 8-inch/DN200 Size Only: Pry the edge of the old seal
washer from inside the clapper seal, as shown above.
5. For the 8-inch/DN200 Size Only: Remove and discard the old
seal washer.
6. Pry the old clapper seal out of the clapper. For the 8-inch/DN200
size, verify that the seal ring is removed with the clapper seal.
Discard the old clapper seal and replace it with a new, Victaulic-
supplied clapper seal assembly. Proceed to step 6a for the 8-inch/
DN200 size or step 7 for all other sizes.
I-751.Europe_22REV_G

Other manuals for Victaulic FireLock 751 Series

Related product manuals