from 18 to 28 ℃ ,within one year after calibration)
N ote 1: without accessory lead resistance
N ote 2: range of exciting current: 0.5mA to 2mA;
max. output voltage: ≤ 2V
N ote 3: temperature coefficient : ± 0.005 % of range per ℃ for the
temperature ranges from 0 to 18 ℃ and from 28 to 50 ℃
G
G
G
G
eneral
eneral
eneral
eneral
Specifications
Specifications
Specifications
Specifications
P ower supply: single 9v battery (ANSI/NFDA 1604A
or
IEC6LR619V alkaline)
B attery life: ca.25 hours
M ax. permitted voltage: 30v (between any two terminals
or
between any terminal and earth
ground)
O perating temperature: 0 ℃ to 50 ℃
O perating relative humidity: ≤ 80 ﹪ RH
S torage temperature: ≤ -10 ℃ to 55 ℃
S torage humidity: ≤ 90 ﹪ RH
S ize: 200 × 100 × 40 mm(with holster)
W eight: 550g (with holster)
Input
Input
Input
Input
Range
Range
Range
Range
Resolution
Resolution
Resolution
Resolution
Accuracy
Accuracy
Accuracy
Accuracy
± 0.05 % of
means.
value ±
0.2
Ω
± 1mA means.
current Notes
1 & 2
± 0.05 % of
set value ±
0.6 ℃
± 1mA exciting
current, use
Pt100-385.
measuring current
1mA Notes 1 & 2
± 1mA exciting
current
Notes 1 & 2