IFU-01, Rev. A Page 11 of 20
Guidance and declaration – electromagnetic immunity
(For ME equipment ME system that are not life-supporting)
Vital Connect Sensor is intended for use in the electromagnetic environment specified below.
The end user of the Vital Connect Platform (including Vital Connect Sensor) should assure that
it is used in such an environment.
Electromagnetic environment- guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.