EasyManua.ls Logo

VitalConnect Platform - Data Accuracy; Error Codes; Description; Service Provider

VitalConnect Platform
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IFU-01, Rev. A Page 7 of 20
Recommendations:
For a good connection and proper operation, the VitalConnect Sensor should NOT be
worm over areas with a high concentration of body hair in the torso area. Removal of
body hair on the torso area is recommended before placing the VitalConnect Sensor.
Recommend gently wiping the sensor module with a damp cloth, after every use
Recommended operating temperature range: 10 - 43 deg C
Recommended shipping and storage temperature range: 0 50 deg C
Recommended shipping and storage relative humidity range: 10 95% RH
Recommended operating relative humidity range: 10 95% RH
Recommended operating altitude: < 2000 m
Recommended barometric pressure: 80 kPa
Data Accuracy:
Data accuracy for the following attributes is measured through various bench testing test cases
in compliance with ANSI/AAMI/IEC 60601-2-47: 2012. They include:
clause 201.12.4.4.101, Linearity and Dynamic Range,
clause 201.12.4.4.102, Input Impedance
clause 201.12.4.4.103, Common Mode Rejection
clause 201.12.4.4.104, Gain Accuracy
clause 201.12.4.4.105, Gain Stability
clause 201.12.4.4.106, System Noise
clause 201.12.4.4.108, Frequency Response
clause 201.12.4.4.110, Timing Accuracy
clause 201.12.1.101.3.1, Heart Rate
clause 201.12.1.101.3.7, ECG Hard Copy
Error Codes/ Description:
VitalConnect Platform is capable of generating the following user error message:
Communication not established
Communication lost
Battery low
Bad battery
Service Provider:

Related product manuals