Vitalograph COMPACT - Instructions for Use
8.2. Product Useful Life Checks
To ascertain whether the device has exceeded its useful life Vitalograph
recommendscheckingtheowheadandtheRealTimeClockbattery.
Theowheadmaybecheckedwiththerecommendeddailycalibration
verication,tobecompletedbytheclinician/healthcareprofessional,and
during the periodic inspection of the device. Reference the calibration
vericationsectionintheSpirotracInstructionsforUse,fordetailsonhowto
checkthedeviceowhead.
The clock may reset to a default time and date on power down if the 3V coin
cell battery has depleted. Vitalograph recommend changing the battery during
routine service every 5 years.
9. Customer Service
Service and repairs should be carried out only by the manufacturer, or by
Service Agents approved by Vitalograph. Contact information for approved
Vitalograph Service Agents may be found at the start of this manual.
Any serious incident that has occurred in relation to the device should be
reported to Vitalograph or its Authorised Representative and the Regulatory
Authorities of the country. Refer to the Vitalograph contact information at the
start of this manual.
It is recommended the coin cell battery powering the RTC is changed every 5
years by a trained Vitalograph service agent or the manufacturer.
10. Consumables and Accessories
Type Cat. No Description
Consumables 28572 Eco BVF and Disposable Nose Clip (80)
28370 28350 BV Filter + 20243 Nose Clips - box of 50
32956 SNIP Nasal Probe (pack of 30, 10 x small, 10 x
medium, 10 x large)
Accessories 36020 2040 Precision Syringe 3-L
67181 COMPACT 19V Power Supply
77933 Flow Conditioning Mesh (10)
77934 Flowhead Complete
77939 Flowhead Cone
77938 Flowhead End Cap
79192 Flowhead Connection Tube
2120013 O-Ring (15)