6Page
ENG
DT_0006 Issue 17
A Global Leader in Respiratory Solutions
of 28
cm(12inches)toanypartoftheCompact,includingcablesspeciedby
Vitalograph. Otherwise, degradation of the performance of this equipment
could result.
22. Use of this equipment adjacent to or stacked with other equipment should
be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to
verify that they are operating normally.
23. TheappliedpartistheFleischowhead.ThisalongwiththeBVF,arethe
contact points for the subject during a spirometry session.
24. Only use the Vitalograph Compact with the medically approved low voltage
power supply with which it is supplied. Attempted use with other power
sources may cause irreparable damage and invalidate the warranty. The
output from the power supply is 19 volts DC.
25. The device may be susceptible when exposed to high levels of
electrostaticdischarge.Torestorefunction,presstheOn/Offswitch.
26. To isolate the Model 6600 Compact from the mains supply, remove the 19V
Power Supply from the mains wall socket. Do not position the 19V Power
Supplywhereitisdiculttounplugfromthesocket.
27. Avoid exposure to known sources of EMI (electromagnetic interference)
such as diathermy, lithotripsy, electrocautery, RFID (Radio Frequency
Identication),andelectromagneticsecuritysystemssuchasanti-theft/
electronic article surveillance systems, metal detectors. Note that the
presence of RFID devices may not be obvious.
28. Loss or degraded performance due to EMI that exceeds the test levels in
immunitytesttablesbelowwillresultinafailedcalibrationverication.
ReferenceSpirotracIFUfordetailsoncompletingacalibrationverication
on the device. If interference is suspected or possible, move the device to a
newlocationandrepeatthecalibrationverication.