washed in warm soapy water and then rinsed in clean water. The
device should be cleaned before and after an extended period of
storage.
5.2. Preventing Cross-Contamination of Subjects in Clinic use
For multi-patient use in a clinic or telemedicine environment
Vitalograph recommends the use of Eco BVF lters or, if these aren’t
available then SafeTway mouthpieces may be used based on the
customer own risk assessment and hygiene controls.
Before use by the next subject, the mouthpiece, outside surfaces
and owhead of the device should be cleaned with a cloth
impregnated with 70% isopropyl alcohol. The device should be
cleaned before and after an extended period of storage. If you
suspect that a device intended for multi-patient use has become
contaminated, it should be replaced.
When used in the clinic environment, it is recommended that
the device be replaced annually. There is no planned preventive
maintenance for this medical device.
6. Fault Finding Guide
Problem Fault
Symptoms:
No ow measurements
Possible
Solutions: (In
probable order)
The batteries may be low. Replace the
batteries.
The owhead may be damaged. Check that
the rotating vane is spinning freely.
Problem Fault
Symptoms:
Cannot read user interface
Possible
Solution:
The batteries may be low. Replace the
batteries.
7. Customer Service
For further assistance, setting up, using or maintaining the device
or to report unexpected operations or changes in performance,
contact Vitalograph, using the contact information at the start of
this manual. Also contact the healthcare provider on any changes to
the performance of the device, as a precaution.