2,000 V 4,000 V ± (1,8% + 4 dgt) 0,001V
20,00 V 40,00 V ± (1,8% + 4 dgt) 0,01V
200,0 V 250,0 V ± (1,8% + 4 dgt) 0,1V
250 V - ± (1,8% + 8 dgt) 1V
"Overload protection 250 V; input resistance: 10 MΩ"
Type of operation DC A (only VC-86)
Range VC-82 Range VC-86 Accuracy Resolution
- 40,00 mA ± (2,0% + 2 dgt) 0,01 mA
- 400,0 mA ± (1,5% + 2 dgt) 0,1 mA
Overload protection 400mA 250 V fast blow fuse
Type of operation AC A 40 to 400 Hz (only VC-86)
Type of operation AC A 40 to 400 Hz (only VC-86)
Range VC-82 Range VC-86 Accuracy Resolution
- 40,00 mA ± (2,2% + 3 dgt) 0,01 mA
- 400,0 mA ± (2,5% + 3 dgt) 0,1 mA
Overload protection 400mA 250 V fast blow fuse
Type of operation Resistance
Range VC-82 Range VC-86 Accuracy Resolution
200,0 Ω 400,0 Ω ± (1,2% + 12 dgt) 0,1 Ω
2,000 kΩ 4,000 kΩ ± (2,0% + 5 dgt) 0,001 kΩ
20,00 kΩ 40,00 kΩ ± (2,0% + 5 dgt) 0,01 kΩ
200,0 Ω 400,0 kΩ ± (2,0% + 5 dgt) 0,1 kΩ
2,000 MΩ 4,000 MΩ ± (3,0% + 2 dgt) 0,001 MΩ
20,00 MΩ 40,00 MΩ ± (5,0% + 2 dgt) 0,01 MΩ
Overload protection 500 V
Continuity tester: acoustic signal for resistances < 35 Ω
"Diode test: Testing voltage max. 3,4V; Display ""OL"" > 2 V"
Attention!
In no event exceed the max. permitted input volu-
mes. Do not contact circuits or parts of circuits if
there could be higher voltages pending within them
than 25 V ACrms or 35 V DC! Danger to life!
41