- 5 -
2.3. PLACE OF USE AND ASSEMBLY
• The balance should be stored and used in locations free of vibrations and shakes, free of
air movement and dust.
• Ambient air temperature should not exceed the range of: +10 °C ÷ +40 °C.
• Ambient relative humidity should not exceed 80%.
• During balance operation, ambient temperature in the weighing room should not change
rapidly.
• The balance should be located on a stable wall console desk or a stable working table
which is not affected by vibrations and distant from heat sources.
• Take special precaution when weighing magnetic objects, as part of the balance is a
strong magnet. Should such loads be weighed, use under-pan weighing option, which
removes the weighed load from area influenced by the balance’s magnet. The hook for
under-pan weighing is installed in balance’s base.
• Keep all package element should your device be transported in the future. Remember that
only original packaging can be used for shipping purposes. Prior to packing, uncouple any
cables, remove any separable components (weighing pan, shields, inserts). Pack the
device components into an original packaging. The original packaging protects the
equipment against potential damage during transportation.
2.4. BALANCE ASSEMBLY
Model: VWR-64B2, VWR-124B2, VWR-214B2
Components-
Installation-
1) Remove the transport lock– gently press the transport lock and turn it accordingly to
<OPEN>, keep the transport lock should your balance be transported in the future.