OLP-5, OLP-6, OLP-8 4-5
4.3 Setting the wavelength
Set the calibrated wavelength of the OLP to the
wavelength of the source signal. You can select
from four or five possible wavelengths. The
wavelengths provided in the power meters
depend on the intended application area:
OLP-5: 820 / 850 / 1300 / 1550 nm
OLP-6: 780 / 850 / 1300 / 1310 / 1550 nm
OLP-8: 950 / 1310 / 1480 / 1550 nm
Note: OLP-5 devices in the A series have the
following calibrated wavelengths:
780 / 850 / 1300 / 1550 nm.
Automatic wavelength detection
In addition to fixed settings of the above
wavelengths, the devices also have automatic
wavelength detection. If this is activated, the “AU”
indicator appears in the modulation field. In this
mode, only TWINtest or Auto-λ signals from the
OLS-5, OLS-6 or OLS-15 system level sources
are detected. If no suitable signal is present, only
dashes appear in the display in the level and
wavelength field.
If a specific wavelength is set, automatic
wavelength detection is deactivated. (Detection
of modulation frequencies of 270 Hz, 330 Hz,
1 kHz and 2 kHz is unaffected by this.)