6-2 OLP-5, OLP-6, OLP-8
Power level measurement
The optical power of unmodulated light (or the
average optical power of modulated light)
emanating from a fiber optical cable is measured
and displayed.
6.2 OLP-6
Wavelength
Automatic wavelength detection “AU”
In combination with the OLS-5, OLS-6, OLS-15
Display range -60 to +5 dBm
Max. allowable power level +10 dBm
Intrinsic meas. error
1
±0.13 dB (±3%)
Linearity
1
(+5 to -55 dBm) ±0.06 dB
1 Under the following reference conditions: -20 dBm (CW),
1310 nm ±1 nm, 23 °C ±3 °C, 45 to 75% rel. humidity
Application range 780 to 1700 nm
Selectable
wavelengths 780/850/1300/
1310/1550 nm
for wavelengths of 780, 850, 1300,
1310, 1550nm
in power level range
780, 850 nm -45 to +10 dBm
1300, 1310, 1550 nm -50 to +10 dBm
Automatic wavelength detection can be switched
off