Questions and Answers

  • A
    Aaron BarrySep 18, 2025
    Why are no cycles stored in the cycle history menu of my W&H Laboratory Equipment?
    • P
      pamela69Sep 19, 2025
      No cycles may be stored if an electronic board was replaced by service. The memory of the old board cannot be restored. Save periodically the history on the USB pen drive.
  • C
    Craig BakerAug 25, 2025
    What to do if W&H Laboratory Equipment is leaking water at the front?
    • P
      priceglenAug 25, 2025
      If you see water leaking at the front of the W&H sterilizer, it could be due to leaks through the chamber door seal. Try cleaning or replacing the door seal and cleaning the chamber face side. If the leak persists, it might be an internal leak, and you should call technical service.
  • S
    Sheryl DavisAug 25, 2025
    Why does W&H Laboratory Equipment chamber door lock and reopen immediately?
    • T
      Troy OnealAug 26, 2025
      If, when starting a cycle, the chamber door locks but re-opens immediately and the 'Open the door' message appears on your W&H Laboratory Equipment, make sure the door seal is evenly inserted on the entire circumference. Also, remove any objects interfering with the chamber door and ensure the door isn't forced against the load or chamber furniture.
  • E
    Eddie MortonAug 27, 2025
    How to prevent corrosion on instruments in W&H Laboratory Equipment?
    • M
      Monique BellAug 27, 2025
      If instruments are showing corrosion or spots after sterilization in your W&H Laboratory Equipment, ensure that instruments are dry before placing them in the sterilizer. Drain both water tanks and use water of good quality. Clean, rinse, and dry instruments to remove any organic or chemical residues before sterilization. Clean the chamber, wash the chamber furniture, and ensure that instruments of different materials do not touch. Also, clean the chamber to remove any scale deposits.
  • E
    Elizabeth SawyerSep 23, 2025
    What does it mean if the W&H Laboratory Equipment display reads “Open the door” but opening the door is impossible?
    • J
      Joseph RichardsonSep 23, 2025
      If the display reads “Open the door” but you can't open it, the chamber might be in a vacuum due to an internal malfunction, or the bacteriological filter is blocked. Switch the sterilizer OFF: this releases any internal pressures allowing the chamber door to be opened. If that doesn't work, remove the bacteriological filter to release the pressure and replace the filter.
  • G
    gutierreztiffanySep 23, 2025
    What does it mean if there is a warning about USB saving (HTML and SCL files) on my W&H Laboratory Equipment?
    • J
      Johnny WhiteSep 23, 2025
      A warning about USB saving typically means the USB pen drive is either not connected or not properly connected to the sterilizer. Check the presence and connection of the USB pen drive.
  • J
    James GoodwinSep 10, 2025
    What to do if instruments are turning brown or black in my W&H Lisa VA131-22?
    • C
      Cindy CobbSep 10, 2025
      If instruments are turning brown or black, it could be due to an incorrect temperature setting. Select a sterilization cycle featuring a lower sterilization temperature, following the instrument manufacturer's instructions.
  • H
    Hannah SmithSep 12, 2025
    What to do if the W&H Lisa VA131-22 cycle report printer does not work?
    • R
      rayjohnSep 12, 2025
      If the cycle report printer isn't working, check the data and power connections to the printer.
  • J
    Janet SmithSep 3, 2025
    Why is W&H Laboratory Equipment not filling water automatically?
    • C
      collierhannahSep 4, 2025
      If, when the sterilizer is connected to an automated water supply system, there is no clean water in the tank, but the automatic water filling does not fill the water in your W&H Laboratory Equipment, connect the water fill system to the sterilizer. If water was temporarily unavailable when the water fill system attempted to fill the tank, switch the sterilizer OFF and then ON again, check the external water supply system, and check for water leaks from the sterilizer.
  • M
    Michael TranSep 1, 2025
    How to fix W&H Lisa VA131-22 cycle report printer?
    • A
      amandamillerSep 2, 2025
      If the cycle report printer does not work with your W&H Laboratory Equipment, check the data and the power connection to the printer.

Summary

Safety and Responsibility

Safety Warnings and Hazards

Highlights critical safety warnings related to thermal, electrical, and operational risks.

User and Manufacturer Responsibility

Defines the responsibilities of the user and the manufacturer for safe operation.

Getting Started

Installation and Connections

Guides on location requirements, electrical, water, and network connections.

Operation and User Interface

Explains basic operation, power control, and homepage elements.

Administrator Functions

User Management and Security

Covers adding, deleting users, and managing PINs for administrative control.

Printer Management

Sterilizer Tests and Validation

Performance Tests (Bowie;Dick, Helix, Vacuum)

Instructions for conducting Bowie/Dick, Helix, and Vacuum tests for performance validation.

Sterilization Cycles

Cycle Operation and Management

Guides on running, scheduling, stopping, and managing sterilization cycles.

Technical Data and Specifications

Water Quality and Validation

Details feed water specifications and validation recommendations.