PGT
®
120.COM Art.Nr. 7100.PGT120.COM
9 / 17 2016-09-07
5 Configuration
The unit can be configured with the DIP switches on the rear according to the table
below.
Standard settings are marked bold.
Switch 1
Switch 2
Test mode
OFF
OFF
"OR" (wrist strap or footwear test)
ON
OFF
Only footwear test
OFF
ON
Only wrist strap test
ON
ON
"AND" (wrist strap and footwear test)
Switch 3
Switch 4
Footwear upper limit
OFF
OFF
20 M for single shoe / 40 M for series
OFF
ON
35 M for single shoe / 70 M for series
ON
OFF
70 M for single shoe / 140 M for series
ON
ON
100 M for single shoe / 200 M for series
Switch 5
Lower limit
OFF
Lower limit disabled
ON
Lower limit enabled
Switch 6
Switch 7
Test voltage
OFF
OFF
30 V
OFF
ON
50 V
ON
--
100 V
Switch 8
Footwear test mode
OFF
test according to switch 1 + 2
ON
footwear in series active
Switch 9
Piep for footwear in series
OFF
At start of test
ON
At end of test if shoes OK
Switch 10
Door opener time
OFF
3s
ON
1s
Switch 11
Not used
OFF
-
ON
-
Switch 12
beeper
OFF
inactive
ON
active