The WaveSense Jazz System has 2 levels of control solution:
Normal and High. Select the level of control solution you
want to use.
The control solution is used to:
[1]
Ensure that your WaveSense Jazz Meter and Test Strips are
working together properly.
[2] Practice testing without having to use your own blood.
Control solution tests should be performed when you:
[1]
First get your WaveSense Jazz Meter.
[2] Begin using a new vial of WaveSense Jazz Test Strips.
[3] Suspect that your WaveSense Jazz Meter or Test Strips are
not working properly.
[4] Think your test results are not accurate.
[5] Have dropped or damaged your meter or exposed your
meter to liquids other than blood samples.
[6] Are advised by your healthcare professional to do so.
Performing a Control Solution Test
7
CHAPTER
[1] Insert a WaveSense Jazz Test Strip: Insert a new
WaveSense Jazz Test Strip into the
meter. Make sure you insert the test
strip, with the contact bars facing up,
into the strip port. The meter will be
activated automatically.
Note: If you do not start the test
within 5 minutes, the meter will be deactivated. To reactivate
your meter, press the CHECK (
) button or remove the unused
WaveSense Jazz Test Strip and reinsert it into the meter.
[2] Meter Ready for Test: The meter now
displays the blood drop symbol and the word
“Fill”. The WaveSense Jazz System is now ready
for you to apply control solution or blood.
Preparing For a Test With Your
WaveSense Jazz Meter
6
CHAPTER

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the WaveSense jazz and is the answer not in the manual?

WaveSense jazz Specifications

General IconGeneral
Sample Size0.5 µL
Test Time5 seconds
CodingNo Coding Required
Operating Temperature50-104°F (10-40°C)
Operating Humidity10-90% RH
Hematocrit Range30-55%
Measurement Range20-600 mg/dL
Storage Conditions36-86°F (2-30°C)