EasyManua.ls Logo

WaveSense jazz - Final Steps

WaveSense jazz
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
52 53
Battery Icon:
Battery #2 has died during test.
Replace battery #2 immediately before retesting.
Troubleshooting Situation #1: Meter does not enter the TEST
Mode after inserting a WaveSense Jazz Test Strip.
[1] CAUSE: The batteries have insufficient power. ACTION:
Replace the batteries, set time and date, and test.
[2] CAUSE: The batteries are installed incorrectly or there are no
batteries in the meter. ACTION: Check that both batteries are
installed correctly, with the positive (+) sign facing up towards you.
[3] CAUSE: WaveSense Jazz Test Strip has been inserted upside
down, wrong end in, or incompletely inserted into the meter.
ACTION: Insert the WaveSense Jazz Test Strip with the black side
up and the contact bars of the test strip in the meter. Ensure that
the test strip is fully inserted.
[4] CAUSE: Defective meter or defective WaveSense Jazz Test
Strips. ACTION: Call customer service.
[5] CAUSE: Blood or foreign objects introduced into the
WaveSense Jazz Meter strip port. ACTION: Please refer to
chapter 22. If problem persists, call customer service and have
the meter serial number available (located on back of the meter).
Troubleshooting Situation #2: The glucose test sequence
does not start after applying the blood sample.
[1] CAUSE: Defective test strip. ACTION: Repeat the test
with a new WaveSense Jazz Test Strip. If problem persists,
call customer service and have the test strip LOT number
Error 5:
The meter was unable to produce a reliable result.
[1] This may be caused by a non-blood sample, or a combina-
tion of high glucose and other medical conditions.
[2] If this error code persists on retesting, consult your health-
care professional.
Error 6:
The test strip has taken too long to generate a signal.
[1] This may be caused by a combination of cold operating
temperature and high hematocrit levels. Retest in a warmer
location.
[2] If this error code persists on retesting, consult your
healthcare professional.
Error 7:
Meter problems have occurred that are beyond your
control.
Contact customer service for assistance.
Temp Icon:
Test was performed outside the system operating
temperature range of 50 °F to 104 °F (10 °C to 40 °C).
Move to an area with an ambient temperature of 50 °F to 104
°F (10 °C to 40 °C). Wait for the meter and test strips to reach
the new temperature (usually 10-20 minutes) and retest.