EasyManua.ls Logo

Wavetek 142 - Page 24

Wavetek 142
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13.
With
FREQ
DIAL
still fully clockwise,
set the voltage 26. Adjust R165
for
a 2.5
V
p-p
signal
output
from
the
between IC2(2)
and
ground (at R95) to 0
Vdc sine amplifier
±25 mV.
±100
/iV,
using
R93.
14.
Set
FREQ DIAL to
1
and adjust
voltage between
TP6 and TP7 to
0
Vdc ±200
juV, using R129.
15. Adjust
voltage
between TPS and
TP9
to 0
Vdc
±200
juV, using
R 1 32.
4.2.5
Triangle Amplitude
Adjustment
16.
Set
FREQ HZ to
X
100,
FREQ
DIAL to
10,
WAVE-
FQRM SELECTOR to
TRIANGLE,
and
SYMMETRY
to
NORM.
17.
Connect the
oscilloscope probe (a
type W or 1A5
plug-in module
should be used
with the scope) to
\
position
of the
WAVEFORM selector
switch,
SW2,
see
Figure
4-4.
18.
Adjust
the triangle
amplitude to
±1.25 V
±5 mV,
using
R12 and R23,
see
Drawing
142-010.
4.2.6
Sine
Distortion
Figure
4-1
-
Distortion Analysis Test
Set-Up
19. Set
FREQ HZ
to
X
100,
VERNIER to CAL, WAVE-
FORM SELECTOR to
SINE,
FREQ
DIAL to
10,
and
OUTPUT ATTEN to 0 dB.
20.
Set 30 V
P-P
MAX
control to 3/4ths of full
CW,
and connect oscilloscope
probe
to
position of
the
WAVEFORM selector switch (SW2), see Figure
4-4.
27. Readjust R153 so that
the
dc
offset
of the sine
amplifier
is 0 V
±5
mV.
28. Connect the Spectrum
Analyzer
as
shown
in Figure
4-2, and
check the sine wave distortion at frequencies
higher than
600
kHz. If distortion
does
not
meet
the
specification, check calibration of
C95,
follow proce-
dure 4.2.10.
21. Adjust
R153 so
that the
dc
offset of the sine ampli-
fier is
zero.
22.
Connect the test equipment as
shown in
Figure
4-1,
and
adjust
R159, R168, and
R145 for
minimum sine
wave distortion.
23. Set
FREQ HZ to X
IK and readjust R159, R168,
and R145
for optimum
sine distortion.
24. Set
FREQ HZ to X 100
and readjust R159, R168,
and R145
if necessary
25. Alternately
perform steps
24 and 25 until the least
amount
of sine wave distortion is obtained
on
both
ranges
typically
.14%.
4.2.7
Frequency Calibration
29.
Set
FREQ
HZ to
X10K, FREQ
DIAL to
10,
VERNIER to
CAL, and
adjust
R269
to
obtain
100 kHz ±100 Hz.
MODEL 142
HP
141S
50-OHM
TERM.
50J2OUTO-
-O
INPUT
Figure
4-2
-
Spectrum Analysis Test Set-Up
4-3
12/71

Related product manuals