11.9 Emission of electromagnetic interference
Guidelines and manufacturer declaration - emission of electromagnetic
interference
The VENTIlogic LS/VENTIlogic plus is intended for operation in an ELECTROMAGNETIC
ENVIRONMENT as outlined below. The customer or the user of the VENTIlogic LS/
VENTIlogic plus should ensure that it is operated in an environment of this type.
Measurements of
interference
emission
Compliance ELECTROMAGNETIC ENVIRONMENT guideline
HF emissions to CISPR
11
Group 1
The VENTIlogic LS/VENTIlogic plus uses HF energy
exclusively for its internal FUNCTION. As a
consequence, its HF emission is very low and it is
unlikely to interfere with adjacent electronic
devices.
HF emissions to CISPR
11
Class B
The VENTIlogic LS/VENTIlogic plus is suitable for use
in facilities other than the domestic sphere and in
those which are connected directly to the PUBLIC
POWER SUPPLY NETWORK which also supplies
buildings used for habitation purposes as long as
the warning below is observed.
Warning: this device/system is intended only for use
by medical specialists. This is a Class A device/
system to CISPR 11. In the domestic sphere, this
device/system may cause radio interference with
the result that in this case, it may be necessary to
take suitable remedial measures, such as realigning,
rearranging or shielding the VENTIlogic LS/
VENTIlogic plus, for example, or filtering the
connection to the premises.
Emission of oscillations
IEC 61000-3-2
Class A
Emission of voltage
fluctuations/flicker to
IEC 61000-3-3
Complies