EasyManuals Logo

Weinmann VENTIlogic LS User Manual

Weinmann VENTIlogic LS
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
84 Troubleshooting
EN
The default display appears again after the acoustic
alarm has been acknowledged. The fault which has not
yet been rectified continues to be displayed in the status
line and the alarm LED flashes (or stays on) until the fault
is rectified.
If the fault is not rectified within 120 seconds of the
acknowledgement, the acoustic alarm (buzzer) sounds
again.
You can find troubleshooting assistance in the following tables.
8.2.3 Sequence of displays if alarms are triggered simultaneously
If several alarms are triggered simultaneously, they are displayed in accordance with the
hierarchy shown below.
1.High-priority alarms
2.Medium-priority alarms
3.Low-priority alarms
If a new alarm with a higher priority occurs after one alarm is displayed, the alarm display
switches to the higher priority. The low-priority alarm is retained and is displayed again
once the high-priority alarm has been eliminated if it still applies.
8.2.4 Physiological alarms
Display Alarm Cause of fault Remedy
IPAP
low
(pressure-controlled
only)
Minimum therapy
pressure undershot.
Medium priority
Filter dirty Clean/change filter
Patient/ventilator
interface leaking
Adjust headgear/headband so
that the patient/ventilator
interface seals, possibly
replace it
Patient/ventilator
interface defective
Replace patient/ventilator
interface
Settings implausible
Have the settings checked by
the doctor supervising
treatment

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann VENTIlogic LS and is the answer not in the manual?

Weinmann VENTIlogic LS Specifications

General IconGeneral
BrandWeinmann
ModelVENTIlogic LS
CategoryMedical Equipment
LanguageEnglish

Related product manuals