3. Maintenance
Clean the equipment
Keep the PRO ECG Recorder, reusable electrodes, and the patient
cable clean. Patient contact with contaminated equipment can
spread infection.
The patient cables should be cleaned after each use and disinfected
as needed.
Never immerse the PRO ECG Recorder or the patient cable in
liquid. Never autoclave or steam clean the PRO ECG Recorder or
the patient cable. Never pour alcohol directly on the PRO ECG
Recorder or the patient cable, and never soak any components in
alcohol. If any liquid enters the PRO ECG Recorder, remove the
PRO ECG Recorder from service, and have it inspected by a
qualified service person before using it again.
Clean the PRO ECG Recorder
Unless the PRO ECG Recorder (the device itself rather than the cables and electrodes)
comes in direct contact with a patient, it is acceptable to clean the device on a routine basis
according to your facility's protocols and standards or local regulations. Patient contact with
the device might require immediate cleaning.
The following agents are compatible with the PRO ECG Recorder:
• 70 percent isopropyl alcohol
• 10 percent chlorine bleach solution
When cleaning the device, avoid using cloths or solutions that
include quaternary ammonium compounds (ammonium chlorides)
or glutaraldehyde-based disinfectants.
1. Slightly dampen a soft, clean cloth with an acceptable
cleaning solution, or use a 70 percent alcohol pad to wipe the
exterior of the PRO ECG Recorder. Dry all components with a
clean, soft cloth or paper towel.
2. Before you power on the PRO ECG Recorder again, wait at
least 10 minutes for all traces of liquid to evaporate.