EasyManuals Logo

Welch Allyn SureTemp Plus 690 User Manual

Welch Allyn SureTemp Plus 690
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #5 background imageLoading...
Page #5 background image
SureTemp® Plus Service Manual 1
Electromagnetic Immunity and Emissions Declaration
Immunity Standard
Emissions Summary
The SureTemp
®
Plus model 690 and 692 thermometers are intended for use in the
electromagnetic environment specified below. The customer or the user of the model 690 and
692 should assure that it is used in such an environment.
IEC Basic Standard IEC 60601-1-2 2001
Electrostatic discharge IEC 61000-4-2
± 6kV
c
± 8kV
c
RF EMF Immunity IEC 61000-4-3
3/10 V/m
80 MHz-2.5GHz
Magnetic field IEC 61000-4-8
3A/m
150kHz-80MHz
Guidance and Manufacturer’s declaration - electromagnetic emissions
Emissions test Compliance
Electromagnetic
environment-guidance
RF emissions
CISPR 11
Group 1
Class B
The SureTemp
®
Plus model
690 and 692 Thermometers
are suitable for use in all
establishments, including
establishments that are
connected to the public
low-voltage power supply
network.
Harmonic emissions
IEC 61000-3-2
Complies
Voltage fluctuations
and flicker emissions
IEC 61000-3-3
Complies

Other manuals for Welch Allyn SureTemp Plus 690

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn SureTemp Plus 690 and is the answer not in the manual?

Welch Allyn SureTemp Plus 690 Specifications

General IconGeneral
BrandWelch Allyn
ModelSureTemp Plus 690
CategoryThermometer
LanguageEnglish

Related product manuals