Immunity Test IEC 60601
Test level
Compliance level Electromagnetic
environment
– guidance
Electrical fast
transients/burst
IEC
61000-4-4
± 2 kV for power
line supplies
± 1 kV for input/
output lines
Not
applicable *)
Not applicable *)
Not
applicable *)
Surge
IEC
61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Not
applicable *)
Not
applicable *)
Not
applicable *)