4. Design and function
The case is rotatable by 330°. When rotating do not touch the display.
4.8 Equipotential bonding
The instrument must be included in the equipotential bonding / grounding of the
application via the process connection. The sealing, e.g. for NPT process connection,
must be conductive to avoid potential differences caused by isolated mounting.
4.9 Integrated data logger
The CPG1500 features an integrated data logger.
This data logger can be switched on and set via “Menu” / “Logger” (respectively).
If the log rate time is longer than the measuring rate, the CPG1500 will take the average
pressure value instead of the current pressure value.
Example:
The mean value should be measured over 60 seconds.
▶
Setting log rate: 60 s
▶
Measuring rate: ≥ 50/s
⇒
Every 60 seconds 1x P_ave., 1x P_max, 1+P_min and 1+temp. will be
recorded