3. Specications
When designing the system, please note that the values given (e.g. burst pressure, overpressure limit) are dependent
upon the material and thread used.
3.1 Measuring ranges and overpressure limits (for measuring range see product label)
Gauge pressure in bar
Measuring range 0 ... 0.1 0 ... 0.16 0 ... 0.25 0 ... 0.4 0 ... 0.6 0 ... 1 0 ... 1.6
Overpressure limit 1.4 1.4 1.4 4.1 4.1 4.1 8.3
Measuring range 0 ... 2.5 0 ... 4 0 ... 6 0 ... 10 0 ... 16 0 ... 25 0 ... 40
Overpressure limit 8.3 19.3 41.4 41.4 82.8 82.8 80
Measuring range 0 ... 60 0 ... 100 0 ...160 0 ... 250 0 ... 400 0 ... 600 0 ... 1,000
1)
Overpressure limit 120 200 320 500 800 1,200 1,500
Measuring range 0 ... 1,600
1) 2)
0 ... 2,500
1) 2)
0 ... 4,000
1) 2)
0 ... 5,000
1) 2)
0 ... 6,000
1) 2)
Overpressure limit 2,300 3,500 5,000 6,000 7,000
1) Only for instruments without ush process connection
2) Only for instruments with ignition protection type Ex i. Not for instruments with SIL2.
Absolute pressure in bar
Measuring range 0 ... 0.25 0.8 ... 1.2 0 ... 0.4 0 ... 0.6 0 ... 1 0 ... 1.6 0 ... 2.5
Overpressure limit 1.4 4.1 4.1 4.1 4.1 8.3 8.3
Measuring range 0 ... 4 0 ... 6 0 ... 10 0 ... 16 0 ... 25
Overpressure limit 19.3 41.4 41.4 82.8 82.8