Wincomm WMP-17F - Guidance and Manufacturers Declaration - Electromagnetic Immunity

Wincomm WMP-17F
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WMP-17F User’s manual
8
For transmitters rated at a maximum output power not listed above, the
recommended separation distanced in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Guidance and manufacturer’s declaration electromagnetic
immunity
The model WMP-17F is intended for use in the electromagnetic
environment specified below. The customer or the user of the model
WMP-17F should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least 30
%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for
power
supply lines
1 kV for
input/output
lines
2 kV for
power
supply lines
1 kV for
input/output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV line(s)
to line(s)
2 kV line(s)
to earth
1 kV line(s)
to line(s)
2 kV line(s)
to earth
Mains power quality
should be that of a
typical commercial or
hospital environment.