SafeLift
Mobilisations-Assistent
Mobilization aid
Lassistant de mobilisation
Mobilisatie-assistent
Assistenza alla mobilizzazione
Asistente de movilización
Gebrauchsanweisung
Instruction manual
Notice d‘utilisation
Gebruiksaanwijzing
Istruzioni d´uso
Manual de Instrucciones

Table of Contents

Questions and Answers

  • J
    Jordan ChangAug 6, 2025
    What to do if my Wissner-Bosserhoff SafeLift bed does not perform any movements?
    • C
      Cody JonesAug 6, 2025
      If your Wissner-Bosserhoff Medical Equipment bed isn't moving, several factors could be at play. First, ensure the GO button is activated by pressing it. Check if the receiver is locked and unlock it if necessary. Verify that the SafeLift Box is correctly connected by checking the plug connection. Also, confirm that the handset is properly plugged in. If the issue persists, the controller or SafeLift box may be defective, or the handset itself might be faulty, in which case you should contact WiBo service.
  • M
    Mary AlvaradoAug 14, 2025
    Why my Wissner-Bosserhoff Medical Equipment bed does not perform any movements?
    • L
      Lawrence HughesAug 14, 2025
      Your Wissner-Bosserhoff Medical Equipment bed might not be moving because the GO button isn't activated; press it to resolve. Alternatively, the receiver could be locked, so unlock it. A not programmed controller can also be the cause, in this case program the controller according to chapter 4.7. Ensure the SafeLift Box is correctly connected by checking the plug connection. The handset might not be correctly plugged in; check its connection. If none of these resolve the issue, the controller, SafeLift box, or handset may be defective.
  • N
    Nicole SolisAug 20, 2025
    How to fix Wissner-Bosserhoff Medical Equipment when bed does not perform any actions?
    • M
      Michael HernandezAug 21, 2025
      If your Wissner-Bosserhoff Medical Equipment bed isn't responding, try these steps: 1. Press the GO button to activate it. 2. Unlock the receiver if it's locked. 3. Check the plug connection of the SafeLift Box. 4. Verify the handset is correctly plugged in. If these steps don't work, it may be necessary to contact WiBo service because the controller or SafeLift box could be defective.
  • T
    Todd EdwardsAug 28, 2025
    What to do if Wissner-Bosserhoff Medical Equipment bed does not perform any movements and another controller is programmed?
    • T
      Tammy MckeeAug 28, 2025
      If your Wissner-Bosserhoff Medical Equipment bed isn't moving and another controller in range is programmed in the SafeLift Box, carry out the deletion procedure as described in chapter 4.7 and re-learn the required controller.

Summary

Introduction

About These Instructions for Use

Information about the manual's structure, signs, and symbols used.

Symbols Used in the Text

Explanation of danger, attention, caution symbols, and text formatting.

Intended Purpose and Users

Details SafeLift's classification, intended use, application restrictions, and target users.

Assembly Aid and General Safety Notes

Mentions mounting videos and provides general safety guidelines for product use.

Components, Structure, and Operation

System Components and Structure

Identifies the main physical parts and illustrates the system's connection layout.

Operating Plan and LED Status

Explains controller functions, GO button usage, and status LED indicators.

Setting Up and Installation

Delivery and Initial Checks

Instructions for checking components upon delivery and reporting damage.

Installation on Quadromat 7 Beds

Step-by-step guide for installing SafeLift on beds with Quadromat 7 motors.

Installation on Duomat 9 Beds

Step-by-step guide for installing SafeLift on beds with Duomat 9 motors.

Combination with SafeSense

Details how to install SafeLift when used in conjunction with SafeSense®.

Installation with Bed Battery

Explains installation on beds with batteries and specific motor types.

Attaching Controller to Handrail

Instructions on how to attach and remove the controller from the bed's handrail.

Controller Programming

Steps to connect (program) and disconnect controllers to the SafeLift box.

SafeLift Functions

Backrest and Height Adjustment

How to adjust the patient's backrest and the bed's height using the controller.

Lock;Unlock Function

Instructions for locking or unlocking the controller functions.

Under-Bed Lighting Control

Details how to control the under-bed lighting via the controller and SafeSense®.

Maintenance, Parts, and Disposal

Cleaning Procedures

Procedures and recommended agents for cleaning the SafeLift controller.

Maintenance and Service Recommendations

Information on low maintenance, recommended annual servicing, and warranty conditions.

Spare Parts and Ordering

Information on using original spare parts and how to order them.

Product Disposal Guidelines

Guidelines for disposing of SafeLift components according to WEEE and local regulations.

Technical Information and Troubleshooting

Technical Data and Specifications

Lists technical specifications like input voltage, frequency, protection class, and manufacturer details.

Classification and Applied Standards

Outlines SafeLift's classification as a medical device and relevant standards.

Nameplate Information

Explains the information and symbols found on the product's nameplate.

Transport and Storage Conditions

Specifies ambient conditions for transporting and storing the SafeLift system.

Troubleshooting Common Issues

Provides common errors, their possible causes, and solutions for SafeLift.

Wissner-Bosserhoff SafeLift Specifications