1.3 Measured values
Differential pressure measurements (temperature-compensated piezo
bridge)
Measuring range: ±2 bar
1 Pa resolution in measuring range -125.00 hPa to +125.00 hPa, otherwise 10
Pa
Precision: < 3% of measured value, better than ±6 Pa within a range of ±200
Pa.
Internal temperature measurements (NTC)
Measuring range: -20 °C to 60 °C
Precision: < ±2 °C
Resolution: 0.1 °C
External temperature measurements (optional, air temperature sensor
order no. 9605 or air temperature probe order no. 9611)
Measuring range: -19.9 °C to +99.9 °C
Precision: < ±2 °C
Resolution: 0.1 °C
T98: < 120 s at 1.5 m/s
Humidity measurements (optional, order no.: 7203)
Measuring range: 0% to 100% RH (relative humidity), non-condensing
Precision: < ±2% RH, within a range of 0–90% RH, otherwise < 3% RH
Resolution: 1% RH
1.4 Calculated values
Pressure units – conversion to mbar, hPa, Pa, mm·H
2
0, PSI in accordance
with generally applicable conversion rules
Temperature units – conversion from °C to °F in accordance with generally
applicable conversion rules
Flow rates – Prandtl measurements displayed in m/s, automatic continuous
density correction based on temperature signal
Range: 2–150 m/s
Pipeline volumes automatically from 0.0 to 1000.0 l (tested to 250.0 l)
Leakage rate (0.0 to 300.0 l/h) – l/h in accordance with DVGW-TRG work-
sheet G 624. The physical data needed to convert diverse gas types can be
extracted from a database stored in the device.
Pressure drop – pre-assessments and main assessments in accordance
with DVGW-TRGI worksheet G 600